A portable transcranial magnetic stimulation (TMS) prototype was designed and constructed for use in clinical trials to manage the pain associated with migraine with aura. In this system, a half-sinusoid (optionally a full-sine) TMS pulse is delivered into a circular coil. The pulse triggering is controlled by a switch in the coil handle. The pulse intensity and polarity are adjustable. Measured pulse parameters (period , maximum magnetic field at coil center ) corresponded to expected values and were in the range described for commercially available TMS apparatus. The maximum current density, computed 3 cm below the coil using a simple model of brain anatomy, was approximately 1.9 times the motor threshold, thus ensuring a sufficient margin for reaching the neuronal threshold in the target area.
Skip Nav Destination
e-mail: renaud.charlet@ims-bordeaux.fr
Article navigation
March 2010
Design Innovations
Design and Construction of a Portable Transcranial Magnetic Stimulation (TMS) Apparatus for Migraine Treatment
Renaud Charlet de Sauvage,
e-mail: renaud.charlet@ims-bordeaux.fr
Renaud Charlet de Sauvage
University of Bordeaux 1
, IMS, Site ENSCPB, UMR CNRS 5218, 16, Avenue Pey Berland, 33607, Pessac Cedex, France
Search for other works by this author on:
Anne Beuter,
Anne Beuter
University of Bordeaux 1
, IMS, Site ENSCPB, UMR CNRS 5218, 16, Avenue Pey Berland, 33607, Pessac Cedex, France
Search for other works by this author on:
Isabelle Lagroye,
Isabelle Lagroye
University of Bordeaux 1
, IMS, Site ENSCPB, UMR CNRS 5218, 16, Avenue Pey Berland, 33607, Pessac Cedex, France
Search for other works by this author on:
Bernard Veyret
Bernard Veyret
University of Bordeaux 1
, IMS, Site ENSCPB, UMR CNRS 5218, 16, Avenue Pey Berland, 33607, Pessac Cedex, France
Search for other works by this author on:
Renaud Charlet de Sauvage
University of Bordeaux 1
, IMS, Site ENSCPB, UMR CNRS 5218, 16, Avenue Pey Berland, 33607, Pessac Cedex, Francee-mail: renaud.charlet@ims-bordeaux.fr
Anne Beuter
University of Bordeaux 1
, IMS, Site ENSCPB, UMR CNRS 5218, 16, Avenue Pey Berland, 33607, Pessac Cedex, France
Isabelle Lagroye
University of Bordeaux 1
, IMS, Site ENSCPB, UMR CNRS 5218, 16, Avenue Pey Berland, 33607, Pessac Cedex, France
Bernard Veyret
University of Bordeaux 1
, IMS, Site ENSCPB, UMR CNRS 5218, 16, Avenue Pey Berland, 33607, Pessac Cedex, FranceJ. Med. Devices. Mar 2010, 4(1): 015002 (6 pages)
Published Online: February 4, 2010
Article history
Received:
January 12, 2009
Revised:
April 10, 2009
Online:
February 4, 2010
Published:
February 4, 2010
Citation
Charlet de Sauvage, R., Beuter, A., Lagroye, I., and Veyret, B. (February 4, 2010). "Design and Construction of a Portable Transcranial Magnetic Stimulation (TMS) Apparatus for Migraine Treatment." ASME. J. Med. Devices. March 2010; 4(1): 015002. https://doi.org/10.1115/1.4000616
Download citation file:
Get Email Alerts
Related Articles
High-Density Transcranial Direct Current Stimulation (HD-tDCS): Skin Safety and Comfort
J. Med. Devices (June,2009)
High-Density Transcranial DC Stimulation (HD-tDCS): Targeting Software
J. Med. Devices (June,2009)
High-Density Transcranial Direct Current Stimulation (HD-tDCS): Hardware Interface
J. Med. Devices (June,2009)
Cerebrospinal Fluid Volume Monitoring for Hydrocephalus Therapy
J. Med. Devices (June,2011)
Related Proceedings Papers
Related Chapters
Experimental Studies
Nanoparticles and Brain Tumor Treatment
The Design of a New Bluetooth Adapter
International Conference on Advanced Computer Theory and Engineering, 5th (ICACTE 2012)
Introduction
Nanoparticles and Brain Tumor Treatment